Abstract
A “micro-electrospray” ionization source has been developed that markedly increases the sensitivity of the conventional electrospray source. This was achieved by optimization of the source to accommodate nanoliter flow rates from 300 to 800-nL/min spraying directly from a capillary needle that, for the analysis of peptides, contained C18 liquid chromatography packing as an integrated concentration—desalting device. Thus, a total of 1 fmol of methionine enkephalin was desorbed from the capillary column spray needle, loaded as a 10-μL injection of 100-amol/μL solution. The mass spectrum showed the [M + H] + ion at m/z 574.2 with a signal-to-noise ratio of better than 5:1 from a chromatographic peak with a width of about 12 s. A narrow range (15-u) tandem mass spectrum was obtained for methionine enekphalin from the injection of 500 amol, and a full-scan tandem-mass spectrum was obtained from 50 fmol. For proteins, the average mass measurement accuracy was approximately 100–200 ppm for the injection of 2.5 fmol of apomyoglobin and 20–40 ppm for 200 fmol. Carbonic anhydrase B and bovine serum albumin showed similar mass measurement accuracies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Society for Mass Spectrometry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.