A Geant4-based Monte Carlo model for Heavy-Ion Therapy (MCHIT) is used to study radiation fields of 1H, 4He, 7Li and 12C beams with similar ranges (∼160–180mm) in water. Microdosimetry spectra are simulated for wall-less and walled Tissue Equivalent Proportional Counters (TEPCs) placed outside or inside a phantom, as in experiments performed, respectively, at NIRS, Japan and GSI, Germany. The impact of fragmentation reactions on microdosimetry spectra is investigated for 4He, 7Li and 12C, and contributions from nuclear fragments of different charge are evaluated for various TEPC positions in the phantom. The microdosimetry spectra measured on the beam axis are well described by MCHIT, in particular, in the vicinity of the Bragg peak. However, the simulated spectra for the walled TEPC far from the beam axis are underestimated. Relative Biological Effectiveness (RBE) of the considered beams is estimated using a modified microdosimetric-kinetic model. Calculations show a similar rise of the RBE up to 2.2–2.9 close to the Bragg peak for helium, lithium and carbon beams compared to the modest values of 1–1.2 at the plateau region. Our results suggest that helium and lithium beams are also promising options for cancer therapy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.