Abstract

Experimental techniques have been developed for obtaining single-event microdosimetric spectra from hospital based linear accelerators. Therapeutic electrons of 12, 15, 18, and 20 MeV from Clinic 18 and 20 accelerators have been produced at ultralow dose rates. Details of the experimental methods have been described previously by the authors. Single-event lineal energy spectra for these beams, as measured by a Rossi chamber differ significantly from cobalt-60 both in shape and dose average lineal energy (yd) which, depending on electron energy, can be 20-40% lower. The spectral peaks for electrons are greatly enhanced compared to cobalt. The yd for electrons also differs from previously reported values for 10-15 MV photon beams. These differences are less pronounced than for differences with cobalt. Spectral peaks and shapes correlate well with the actual electron stopping powers in tissue. The theory of dual radiation action predicts changes in relative biological effectiveness at low doses for megavoltage photon and electrons. Although at clinical doses the predicted differences are not statistically significant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.