Abstract

A multi-element tissue-equivalent proportional counter (TEPC), based on a single gas electron multiplier (GEM) foil of standard geometry, has been constructed with 16 cylindrical sensitives volumes. In this article, the design of this novel counter is described and first microdosimetric measurements are presented. To study the response of the GEM-TEPC to both low and high linear energy transfer radiation fields, the microdosimetric spectra due to a 137Cs gamma-ray source and to fast neutrons from 7Li(d,n)8Be reaction have been measured using pure propane gas at low pressure, in order to simulate a tissue site of about 1 µm equivalent size. The comparison with spectra measured with a spherical TEPC and with a mini-TEPC demonstrates promising properties for application of the GEM-TEPC for microdosimetric applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.