Abstract

A microcantilever array biosensor based on a sandwich structure has been developed for simultaneously measuring two biomarkers carcinoembryonic antigen (CEA) and α-fetoprotein (AFP) via an optical readout technique-real-time monitoring of the profile of cantilever. First, the aptamers of CEA and AFP were self-assembled on their respective cantilevers. After the adsorption of the mixture of CEA and AFP, further specific interaction was performed via the addition of the antibodies specific to each target. The compressive stress on the cantilever was generated by the aptamer-antigen-antibody sandwich structure formed on the gold surface, resulting in cantilever bending. The profile of cantilever could be monitored in real time. The relationship between the deflection value at the 90% position of the cantilever and the target concentration served as a calibration curve, and the detection sensitivity was 1.3 ng/mL for CEA and 0.6 ng/mL for AFP, respectively. This work demonstrated the ability of simultaneously measuring two biomarkers via a microcantilever array biosensor, giving great potential for further application in detecting several targets simultaneously for early clinical diagnosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.