Abstract

Microcalorimetric methods have been developed for continuous analytical control of glucose, lactose and sucrose concentrations in solutions in biotechnological processes. The methods are based on the Malprade reaction. The carbohydrate concentrations were detected on a flow-mix microcalorimeter. The accuracy of measurements of the carbohydrate concentration (2–4 mM) in solution is 1–2%. Other components common to biotechnological solutions (amino acids, ammonium ions, urea and acetate) have virtually no influence on the results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.