Abstract

The thermal effect ability of an ultrasound contrast agent (UCA) in high-intensity focused ultrasound (HIFU) therapy was evaluated. This was done using a thermal effect simulation program using a microbubble-cloud model and modified Pennes Bio-Heat Transfer Equation (BHTE). In order to design the proposed simulation program, a spherical microbubble-cloud model was developed. Next, the proposed simulation program was experimentally investigated by ultrasound phantom experiments using commercial HIFU equipment and UCA, and the change of thermal effect according to the microbubble-cloud population ratio was measured. The HIFU thermal effect simulations using the properties of UCA were compared with the phantom experiments, which provided concordant results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.