Abstract

There is currently little evidence regarding potential risks of bacterial contamination of non-invasive ventilation (NIV) devices used by cystic fibrosis (CF) patients. The aim of this study was to determine the extent of bacterial contamination of NIV devices in our regional adult CF centre. Seven NIV devices recently used by CF patients chronically infected with Pseudomonasaeruginosa or Burkholderiacepacia complex (BCC) were swabbed in seven areas, both external and internal. Two devices had undergone ethylene oxide (EtO) sterilization between patient use and swabbing, and five devices had not undergone EtO sterilization. Swabs from five devices had insignificant growth of environmental organisms and two devices had significant growth of environmental organisms. No CF pathogens were isolated from any machine. No evidence was found of pathogenic microbial contamination of NIV devices used by CF patients in this small study. We suggest that further studies examine for evidence of bacterial contamination of NIV devices and that this issue should be included in future CF infection control guidelines.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.