Abstract
To quantify microbial aerosols generated by a series of laboratory accidents and to use these data in risk assessment. A series of laboratory accident scenarios have been devised and the microbial aerosol generated by them has been measured using a range of microbial air samplers. The accident scenarios generating the highest aerosol concentrations were, dropping a fungal plate, dropping a large bottle, centrifuge rotor leaks and a blocked syringe filter. Many of these accidents generated low particle size aerosols, which would be inhaled into the lungs of any exposed laboratory staff. Spray factors (SFs) have been calculated using the results of these experiments as an indicator of the potential for accidents to generate microbial aerosols. Model risk assessments have been described using the SF data. Quantitative risk assessment of laboratory accidents can provide data that can aid the design of containment laboratories and the response to laboratory accidents. A methodology has been described and supporting data provided to allow microbiological safety officers to carry out quantitative risk assessment of laboratory accidents.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.