Abstract

In this 10-min gas-chromatographic assay for theophylline, a 25-microliter portion of serum or saliva sample, 100 microliter of extraction reagent, and 1.0 ml of solvent were mixed, the solvent separated and evaporated, and an aliquot of the reconstituted extract injected into the gas chromatograph. At a concentration of 11 mg/liter, within-run precision (CV) was 3.6% and run-to-run 2.9%. At 22 mg/liter, within-run precision was 2.8%. Concentration and instrument response are linearly related between 2 and 40 mg/liter. The limit of detection was 0.4 mg/liter. Results correlated well (r = 0.98) with those by another micro-scale gas-chromatographic procedure. Analytical recovery was estimated to exceed 90%, and no interferences from other xanthines or related drugs were observed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.