Abstract
Micro-pocket fission detectors (MPFD) have been fabricated and tested as in-core flux monitors in the 250 kW TRIGA nuclear reactor at Kansas State University. The prototype devices have been coated with a natural uranyl-nitrate to provide a neutron reactive coating. The devices are composed of alumina substrates sealed together to form a miniature gas pocket 3 mm in diameter and 1 mm wide. The devices are radiation hard and can operate in pulse mode in a neutron flux exceeding 10 12 cm −2 s −1. Placed in the central thimble of the reactor core, the MPFDs have shown count rate linearity from low to high power. Dead time losses become apparent at power levels exceeding 100 kW, yet are still low enough to allow for pulse mode operation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.