Abstract

A new procedure is described for the visible-range spectrophotometric analysis of glutathione (GSH) in microvolumes of blood (as low as 0.5 μL) collected by fingerstick. Samples are diluted 1 to 300 (v/v) in a stabilizing solution, followed by determination of haemoglobin concentration and by acid deproteination. GSH is then measured in the clear supernatant by colorimetry using DTNB, i.e., 5,5'-dithio-bis(2-nitrobenzoic acid), in aqueous solution at pH 7.8. The DTNB reagent is prepared and kept at pH 6.2 until just prior its addition, thus avoiding spontaneous decomposition of the reagent. The assay is rapid, easy to adapt to large-scale studies and it avoids artefactual oxidation of GSH, a common methodological shortcoming. The method is precise with 1.7 to 3.4% intra-day relative standard deviation (RSD) and 2.2 to 4.2% inter-day RSD, and accurate with -1.4% to 2.3% intra-day relative error (RE) and -2.8% to 1.6% inter-day RE. GSH is recovered by 97.5 to 100% at all tested concentrations. The new colorimetric micro-method was validated by a reliable previously reported HPLC method. The procedure is suitable for minimally invasive investigation of oxidative stress in peripheral blood.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.