Abstract
A new simplified and rapid method for detection and quantitation of "carbohydrate-deficient transferrin" in serum is described. The method is based on isocratic anion exchange chromatography of isotransferrins in disposable microcolumns followed by a double antibody transferrin radioimmune assay. This technique, which separates all transferrin components isoelectric above pH 5.65, showed a very good reproducibility and accuracy with a coefficient of variation between 5 and 9%. 77 alcoholic patients could be clearly separated from 80 healthy "normal consumers" and 33 total abstainers with a specificity of 100% and a sensitivity of 91%. The values were significantly correlated to the amount of alcohol consumed during the latest month, and declined in abstaining alcoholics with a mean biological half-life of 17 days. Elevated levels occasionally appeared in healthy individuals after daily consumption of 60 g of ethanol during a 10-day period. In a sample of 187 patients with nonalcohol-related conditions only 2% false-positive values were found. This method is suggested as a potential tool for detecting and monitoring alcohol abuse.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.