Abstract

Abstract An MeV-SIMS microprobe has been constructed based on primary beam collimation by a conical glass capillary. Molecular surface mapping is obtained by raster scanning of the sample with a piezo-driven XY stage. The angular divergence of the collimated beam is smaller than 0.3 mrad, which allows a relatively large distance between capillary exit and sample without a significant loss in lateral resolution. With a capillary of 5 μm outlet diameter a lateral resolution of approximately 10 μm has been observed at a working distance of 50 mm. Capillary collimation can be obtained in a very simple and inexpensive fashion and is virtually independent of the used beam particle properties over a large range of ion mass and energy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.