Abstract

We designed the interferometric test of a 300 mm flat mirror, based onto a spherical mirror and a dedicated CGH. The spherical beam of the interferometer is quasi collimated to the desired diameter by the spherical mirror, used slightly off-axis, and the CGH performs the residual wavefront correction. We performed tests on a 200 mm and 300 mm flat mirrors, and compared the results to the ones obtained by stitching, showing an accuracy well within the designed value. The possibility to calibrate the cavity by subtracting out the figure errors of the spherical mirror has also been evaluated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.