Abstract

This article proposes a metrological traceability method for polarization extinction ratio (PER) ranging from 0 up to 70 dB, while the common method is limited to 50 dB. A precision alignment coupling point is designed and manufactured using two polarization-maintaining fibers (PMFs). By rotating one of the PMFs to change the axis-alignment angle and to control the distribution of light energy in the fast and slow axis, a traceability chain from the PER to the rotation angle is established. With extremely low-stress clamping of the PMF, 5-D precision adjustment of the fiber position, and precision coaxial rotation in the order of arc seconds, accurate and controllable generation of any PER in the range of 0–70 dB can be achieved. A high PER traceability system is designed and built, which is successfully used in the traceability of PER to the angle of circumference. The expanded uncertainty ( <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$k = 2$ </tex-math></inline-formula> ) in the ranges of 0–30 dB, 30–50 dB, and 50–70 dB is 0.01, 0.06, and 1.20 dB, respectively. Controlled generation of high PER (50–70 dB) with metrological traceability is realized for the first time as far as we know.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.