Abstract

In the framework of the EMPIR research project Metrology for Radon, a facility was devised that provides stable and traceable atmospheres for the realization of radon activity concentration at a high metrological level. The facility is used to calibrate and test devices for measuring radon in the air. It enables precise long-term exposures in the range between 100 Bq/m3 and 1000 Bq/m3. The radon activity concentrations are realized with uncertainties of about 1.0 % (k = 1).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.