Abstract
AbstractAn amperometric glucose bioelectrode based on the incorporation of methylene green (MG) within a glucose oxidase (GOD)‐containing carbon paste is described. The electrode generates an anodic current in the range of 0.0–0.6 V (vs. SCE). The optimum pH is 8.5. The apparent Michaelis constant for the Cali bration plot is 49.3 mM (at 0.5% MG and 5.0% GOD loadings). The electrode sensitivity is 18.1 μA/mM cm2. The electrode is insensitive to paracetamol and ethanol. At 0.068 V the response to ascorbic acid (40 μM) accounts for 2.4% of that of the normal physiological glucose level in blood. The performance was optimized by systematically altering the electrode composition and electrode potential. The electrode was used to measure glucose in a serum sample and gave an acceptable correlation with the reference method Eksan‐G results. Good stability was observed for storage in the dry state in the refrigerator.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.