Abstract

We describe a method for automatic computation of "methylene unit" retention-time indices in real time on a computer-controlled gas chromatograph/mass spectrometer system. The calculation is based on the exponential relationship between the methylene unit values of hydrocarbon reference standards and the oven temperature at their elution. Using this mathematical relationship, we calculated calibration factors, which eliminates the need to include a set of methylene unit reference standards with each run and simplifies and speeds identification of unknown peaks in chromatograms. Long-term stability and good precision of the calibration factors were observed during the six-month period of study.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.