Abstract

GEIGER-MULLER counters filled with methylene bromide and argon are known1 to be self-quenching and to possess long sloping plateaux. Measurements have now been made of the β-counting-rate as a collimated source was moved diametrically across the mica end-window of a β-ray counter so that β-particles entered the counter parallel to the centre wire. Two counters A and B of identical construction were used for these measurements, A containing argon at a pressure of 97 mm. of mercury and ethyl alcohol at a pressure of 9·7 mm. of mercury, and B containing argon at a pressure of 7·4 mm. of mercury and methylene bromide at a pressure of 7·4 mm. of mercury. It was found that the efficiency of counter B was greatest for particles entering the counter near the centre wire, but even then was less than that of counter A by a factor of about three.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.