Abstract
A highly sensitive and rapid electrochemical sensor was developed for detecting melatonin using a molecularly imprinted polymer (MIP) with methylene blue as the functional monomer and melatonin as the template. The MIP was synthesized via a simple electropolymerization process that did not require an initiating reagent. The sensor demonstrated good selectivity for melatonin against common interferences such as lactate, cytosine, cytidine, urea, ascorbic acid, creatine, creatinine, serotonin, and tryptophan. Melatonin detection was achieved ata potential of 0.60V vs. Ag/AgCl with a sensitivity of 138.8 ± 4.7 µA µM‒1 in the linear range 0.097 - 200µM and a limit of detection of 29nM (3SB/m). The sensor exhibited excellent reproducibility and repeatability for both within (intra) and between (inter) electrodes (%RSD < 3% for n = 3). The sensor was applied to authentic urine and saliva samples with recoveries of 103 ± 1% and 102 ± 1%, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.