Abstract

We have demonstrated a novel microbubbles methylene blue solution, called to “MB 2 ” solution for a dual modality contrast. We have photoacoustically and ultrasonically imaged and quantified aqueous solutions of MB 2 by varying the concentration of either microbubbles or methylene blue to investigate the dual modal imaging capability. Interestingly, as the microbubbles concentration increased with the constant methylene blue concentration, photoacoustic (PA) signal was greatly attenuated in the MB 2 solution. Conversely, when methylene blue concentration increased with the fixed microbubbles concentration, no interference was observed in ultrasound (US) signals. To further confirm our findings, we switched the PA and ultrasound (US) signals using conventional ultrasound. We compared the PA and US signals of the MB 2 solution before and after sonication. The PA amplitude increased 2.5 times. Conversely, the US signals were initially strong, but decreased 2.5 times following sonication. Moreover, we used a clinically modified PA/US imaging system to disrupt the microbubbles in MB 2 and recover the PA signals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.