Abstract
To reduce stimulus transduction artifacts in EEG while using insert earphones. Reference Equivalent Threshold SPLs were assessed for Etymotic ER-4B earphones in 15 volunteers. Auditory brainstem responses (ABRs) and middle latency responses (MLRs)-as well as long-duration complex ABRs-to click and /dα/ speech stimuli were recorded in a single-case design. Transduction artifacts occurred in raw EEG responses, but they were eliminated by shielding, counter-phasing (averaging across stimuli 180° out of phase), or rereferencing. Clinical grade ABRs, MLRs, and cABRs can be recorded with a standard digital EEG system and high-fidelity insert earphones, provided one or more techniques are used to remove the stimulus transduction artifact.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.