Abstract

The assessment of the actual measurement accuracy of an intracranial pressure (ICP) transducer is imperative but still very difficult in practice. The authors tested the Codman MicroSensor ICP transducer experimentally. Additionally, a bedside test for assessment of measurement accuracy was used before and after clinical monitoring. For laboratory testing, seven new transducers were examined for measurement accuracy at increasing pressures ranging from 0 to 75 mm Hg. Drift was evaluated for 10 days at six different pressure levels ranging from 0 to 50 mm Hg. Temperature drift was assessed over a temperature range from 20 to 45 degrees C. The percentage absolute difference was used in the assessment of measurement accuracy. For clinical testing, measurement accuracy was assessed intraoperatively just before the insertion of the transducer and in an open water bath after its explantation, at 10 cm H2O both times. The maximum percentage absolute difference was 9% at a pressure of 10 mm Hg, and declined toward 2.3% at a pressure of 75 mm Hg. The maximum drift over 10 days was -4 mm Hg. Within the range of 30 to 40 degrees C, temperature drift was negligible. Over a period of 2 years, between June 1997 and June 1999, 40 ICP transducers were implanted in 35 patients by one surgeon. Of these devices, a malfunction was detected in two of them by testing them in a water bath before insertion. Experimental and clinical results indicate that this miniature strain-gauge transducer measures accurately; however, control readings for the probe by means of measurement in an open water bath just before insertion are strongly recommended.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call