Abstract
An optoelectronic in-line imaging system for particle sizing is described. The images are scanned by two CCD cameras viewing the same object field that is illuminated with two pulses. One of the cameras is double exposed and the other is activated only during the second pulse. Two successive analogical subtractions between the video output signals gives the sign of the transverse velocity vector. Out-of-focus images are deconvolved with the assumption of a Gaussian point-spread function (PSF) whose spatial parameter ς increases with the defocusing distance z. In the case of low particle density, an algorithm based on the exploitation of the power spectral density is used to estimate the particle diameter. This method can be applied to the case of fast-moving particles (e.g., v < 500 m/s). The accuracy of the size measurement is better than 10% in the diameter range 20-160 μm. The main result is that this accuracy is obtained with an amount of defocusing in the range [-2, 2] mm. Thus, the depth of field is significantly extended in comparison with a conventional microscopic imaging system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.