Abstract

Shear force microscopy is not normally associated with the imaging of biomolecules in a liquid environment. Here we show that the recently developed scattered evanescent wave (SEW) detection system, combined with custom-designed vertically oriented cantilevers (VOCs), can reliably produce true non-contact images in liquid of DNA molecules. The range of cantilever spring constants for successful shear force imaging was experimentally identified between 0.05 and 0.09 N m−1. Images of λ-DNA adsorbed on mica in distilled water were obtained at scan rates of 8000 pixels s−1. A new constant-height force mapping mode for VOCs is also presented. This method is shown to control the vertical position of the tip in the sample plane with better than 1 nm accuracy. The force mode is demonstrated by mapping the shear force above λ-DNA molecules adsorbed on mica in a liquid environment at different tip–sample separations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.