Abstract
In this paper, various methods for measuring blood pressure (BP) are considered and discussed in detail. Among those, we focus on the indirect BP measurement using the volume clamp method (VC). We considered a hardware solution using the SAKR-2 (Ltd Intox) device for non-invasive BP measurement. Based on our experimental results, this solution is well suited to BP dynamics measure for the patients both at rest and during movement. Specifically, by analyzing the characteristics of multiple measured results in different scenarios, we have proposed several post-processing techniques to remove the systematic errors during measurements (e.g., due to the measurement condition) and enhance accuracy. Compared to the conventional direct BP measurement methods in the radial artery (RA) using commercial device (S5 monitor), both systolic blood pressure (SBP) and diastolic blood pressure (DBP) indicators have strong relationship. The measured BP shifts with respect to hydrostatic changes in blood pressure are fairly match to the theoretical predetermined value during repetitive measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.