Abstract

Thermal drift of optical systems employed for surface guided radiation therapy (SGRT) adds uncertainty to patient setup and monitoring. This work describes methods to measure the drift of individual camera pods as well as the drift of the combined clinical signal. It presents results for four clinical C-Rad Catalyst+ HD systems. Based on the measured clinical drift, recipes are provided on how to calculate relevant uncertainties in patient setup and patient position monitoring with SGRT. Strategies to reduce the impact of drift are explained. While the results are specific to the systems investigated, the methodology is transferable and the clinical recipes are universally applicable.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.