Abstract
Large-area, multi-pixel photon counters will be used for the electromagnetic Barrel Calorimeter of the GlueX experiment at Jefferson Lab. These photo sensors are based on a $3 \times 3~\hbox{mm}^2$ cell populated by $50~\mu\hbox{m}$ pixels, with 16 such cells tiled in a $4 \times 4$ arrangement in the array. The 16 cells are summed electronically and the signals are amplified. The photon detection efficiency of a group of first-article units at room temperature under conditions similar to those of the experiment was extracted to be ${(28 \pm 2(stat) \pm 2(syst))\% }$ , by employing an analysis methodology based on Poisson statistics carried out on the summed energy signals from the units.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.