Abstract

In this work, a methodology is presented for the determination of the respiration rate of a person under test (PUT), the detection of movements, as well as the elimination of the spurious effects produced by the movements of the PUT. The methodology is based on Empirical Modal Decomposition (EMD) applied to the phase signal obtained by means of a quadrature Doppler radar operating in S band. The EMD allows to automatically eliminate the continuos component (CC) which is present in the phase signal since one of the main characteristics of the modes generated by the EMD is that its mean is equal to zero. On the other hand, the first mode of the EMD is used for the detection of movements while the sum of the second and third modes are used for the elimination of the CC drift caused by the DC drift and the high frequency components produced by the movements of the PUT. The proposed methodology was successfully tested in a PUT at rest and performing movements of the head, arm and combination of head, arm, and torso. The average respiration rate measured was 20.78 breaths / min with a standard deviation of 2.53 breaths/min.