Abstract

Most of the measurement methods for blue light hazard (BLH) evaluation are based on radiance, which are too complex and hard to find out the highest radiance. We proposed a methodology for measurement of BLH with imaging luminance meter. We measured the relative spectral power distribution with a spectrometer and the maximum luminance with an imaging luminance meter for LEDs and compact fluorescent lamps, and calculated the BLH weighted radiance. The BLH efficacy and the upper limit of luminance with blue light safe of various light sources were also calculated on the basis of the BLH weighted function. The results show absolute blue light safe can be obtained if the luminance is limited to 100 kcd m2 for color temperature lower than 6500 K. LEDs with high color temperature, especially those without diffused window, have potential risk of BLH. Our method can be used in online measurement of the BLH of LEDs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.