Flame Atomic Absorption Spectrophotometry (FAAS) has been developed for determination of iron (Fe) in multivitamin mineral capsule dosage form. The method of sample preparation used combination of dry ashing (450 ± 20˚C for 2 hours or until it becomes ashes) and wet ashing with a mixture H2O2 30%, HCl 6N, and HNO3 65%. The method validation comprises of the study of selectivity, linearity, accuracy, and precision. The proposed method was linear over the concentration ranges of 2–8 μg/mL. The regression was y=0, 0435x+0, 0237 with the correlation coefficient (r) obtained for those regression is 0, 9996 and the relative process standard deviation value (Vxo) is 1, 39%. The accuracy of Fe was 100, 12%. The precision coefficients of variation in the range 0, 99% for repeatability and 0, 83% for intermediate precision. The method was successfully applied for determination of iron (Fe) in multivitamin mineral capsule dosage form. The conclusion of this study that the method of FAAS (Flame Atomic Absorption Spectrophotometry) meets the validation requirements in selectivity, linearity, accuracy, and precision.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.