Abstract

A fast, sensitive, and accurate GC/MS method for the quantification of aliphatic nitroesters (ethylene glycol dinitrate, nitroglycerin, and triethylene glycol dinitrate) and aromatic amines (diphenylamine, 2-nitrodiphenylamine, and triphenylamine) in propellants was developed and validated. This method comprises a Soxhlet extraction step with dichloromethane, followed by separation on a capillary column MDN-5. Ionization of the analytes is carried out using electron ionization. The limit of quantification of the method was 1% w/w for aliphatic nitroesters and 0.1% w/w for aromatic amines (diphenylamine and triphenylamine). Values of repeatability and reproducibility for analyzed compounds were smaller than values of the maximum allowed tolerances of the Horwitz-equation RSD(max) and 2/3 RSD(max). Values of accuracy for selected compounds were below the acceptable threshold of 15% for all tested levels in the range of calibration curve excepting the lowest concentration of calibration curve for nitroglycerin and aromatic amines. During the validation of method, temperature instability in injection port of gas chromatograph and column was observed for 2-nitrodiphenylamine. Hence, it follows worse results of accuracy and linearity and 2-nitrodiphenylamine was not validated successfully.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.