Abstract
Results from proficiency tests gathered over the past two decades by the laboratory for low level radioactivity measurements for liquid scintillation counting of 3H (184 results) and 14C (74 results) are used to verify the validated measurement methods used by the laboratory, in particular the estimated uncertainty budget of the method and its reproducibility and stability. A linear regression approach is used for the analysis of the results, described in the literature as the long term uncertainty in measurement method. The present study clearly indicates the advantages of using proficiency test results in identifying possible constant or proportional bias effects as well as the possibility to compare the laboratory performance with the performance of peer laboratories.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.