Abstract

The analytical method for organically bound tritium (OBT) was developed in our laboratory. The optimized operating conditions and parameters were established for sample drying, special combustion, distillation, and measurement on a liquid scintillation spectrometer (LSC). Selected types of OBT samples such as rice, corn, rapeseed, fresh lettuce and pork were analyzed for method validation of recovery rate reproducibility, the minimum detection concentration, and the uncertainty for typical low level environmental sample was evaluated. The combustion water recovery rate of different dried environmental sample was kept at about 80%, the minimum detection concentration of OBT ranged from 0.61 to 0.89 Bq/kg (dry weight), depending on the hydrogen content. It showed that this method is suitable for OBT analysis of environmental sample with stable recovery rate, and the combustion water yield of a sample with weight about 40 g would provide sufficient quantity for measurement on LSC.