Abstract

During transcatheter aortic valve replacement procedures, alignment of the interventional x-ray beam with the aortic valve plane is required to guide valve placement. The purpose of this work was to develop methods to measure x-ray beam angles which are aligned with the aortic valve plane from a single x-ray image of a valvuloplasty balloon inflated across the left ventricular outflow tract (LVOT). A custom valvuloplasty balloon and image analytics were developed. In-vivo pig experiments demonstrated that the angular orientation of the LVOT can be measured with precision less than 3°, including the influence of cardiac motion during rapid pacing. Using these methods, the angular orientation of the LVOT of a large mammal can be measured precisely and the continuous range of x-ray projection angles which are aligned with the aortic valve plane can be calculated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.