Abstract

Abstract Many catheters and vascular implants are coated to increase biocompatibility or to reduce friction during catheter based implantation. Several regulations require assessment of coating durability over the implant’s life time. An in vitro method for stent testing is presented to measure released particulate matter at defined inspection intervals. The method was validated using polystyrene microspheres with a size of 10, 25 and 50 μm to check for particle recovery (n=6). Two cleaning steps followed. Particle counting was performed by light obscuration method. The recovery rate was 103±5% (10μm), 94±4% (25 μm) and 78±12% (50 μm), respectively, meeting the requirements of FDA guidance documents (i.e. FDA 1545). Less than 3% of the particles were found in the cleaning solutions. The method using a fixed volume during stent loading can be adapted to all durability testers where tubes are used to fix the stents (radial pulsatile, bending, axial compression, torsion).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.