Abstract

Here we present a novel method to improve the kinetic energy resolution of a velocity map imaging(VMI) spectrometer. The main modifications, compared to the original design of Eppink and Parker (1997 Rev. Sci. Instrum. 68 3477), are two additional grid electrodes. One of the electrodes is a grounded grid and the other is an arc-shaped grid with negative voltages (or positive voltages for an ions spectrometer). The arc-shaped electrode is axially symmetrical around the spectrometer axis. The field constructed by the two electrodes is to compensate the dispersion of the ‘v’-shaped energy resolution. Simulations by SIMION and reconstructions by the basis set expansion Abel transform method show that the kinetic energy resolution can be improved drastically by our new method. Furthermore, the accuracy in the determination of the kinetic energy of ion/electrons remains unchanged with respect to the original design.