Abstract

BackgroundAn efficient and accurate method of respiratory rate measurement is still missing in hospital general wards and triage. The goal of this study is to propose a method of respiratory rate measurement that has a potential to be used in general wards, triage, and different hospital settings with comparable performance. We propose a method of respiratory rate measurement that combines a unique wearable platform with an adaptive and optical approach. The optical approach is based on a direct-contact optical diffuse reflectance phenomenon. An adaptive algorithm is developed that computes the first respiratory rate and uses it to select a band. The band then chooses a set of unique optimized parameters in the algorithm to calculate and improve the respiratory rate. We developed a study to compare the proposed method against reference manual counts from 82 patients diagnosed with respiratory diseases.ResultsWe found good agreement between the proposed method of respiratory rate measurement and reference manual counts. The performance of the proposed method highlighted deviations with a 95% confidence interval (C.I.) of − 3.34 and 3.67 breaths per minute (bpm) and a mean bias and standard deviation (STD) of 0.05 bpm and 2.56 bpm, respectively.ConclusionsThe performance of the proposed method of respiratory rate measurement is comparable with current manual counting and other respiratory rate devices reported. The method has additional advantages that include ease-of-use, quick setup time, and being mobile for wider clinical use. The proposed method has the potential as a tool to measure respiratory rates in a number of use cases.

Highlights

  • An efficient and accurate method of respiratory rate measurement is still missing in hospital general wards and triage

  • We present a new method of respiratory rate measurement that combines a unique wearable platform for ease of measurement and an accurate and adaptive non-invasive optical approach based on the optical diffuse reflectance phenomenon [14]

  • The Singhealth Centralized Institutional Review Board (IRB) approval was attained for this study (IRB Ref. 2017/2961)

Read more

Summary

Introduction

An efficient and accurate method of respiratory rate measurement is still missing in hospital general wards and triage. The goal of this study is to propose a method of respiratory rate measurement that has a potential to be used in general wards, triage, and different hospital settings with comparable performance. We propose a method of respiratory rate measurement that combines a unique wearable platform with an adaptive and optical approach. An adaptive algorithm is developed that computes the first respiratory rate and uses it to select a band. The band chooses a set of unique optimized parameters in the algorithm to calculate and improve the respiratory rate. We developed a study to compare the proposed method against reference manual counts from 82 patients diagnosed with respiratory diseases. Poor visibility of the start and end of a breath, interruptions, moving patients, difficulty in counting, or remembering a count can lead to further errors [9]

Objectives
Methods
Results
Conclusion