Abstract

A novel technique is presented that addresses the issue of how to apply internal standard (IS) to dried matrix spot (DMS) samples that allows the IS to integrate with the sample prior to extraction. The TouchSpray, a piezo electric spray system, from The Technology Partnership (TTP), was used to apply methanol containing IS to dried blood spot (DBS) samples. It is demonstrated that this method of IS application has the potential to work in practice, for use in quantitative determination of circulating exposures of pharmaceuticals in toxicokinetic and pharmacokinetic studies. Three different methods of IS application were compared: addition of IS to control blood prior to DBS sample preparation (control 1), incorporation into extraction solvent (control 2), and the novel use of TouchSpray technology (test). It is demonstrated that there was no significant difference in accuracy and precision data using these three techniques obtained using both manual extraction and direct elution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.