Abstract

We report here the development of a precise and sensitive method for the determination of vitamin K homologues including phylloquinone (PK), menaquinone-4 (MK-4), and menaquinone-7 (MK-7) in human plasma using HPLC-tandem mass-mass spectrometry with atmospheric pressure chemical ionization (LC-APCI-MS/MS). The method involves the use of stable isotope (18)O-labeled internal standard compounds, which were synthesized in our laboratory, and the selection of a precursor and product ion with a MS/MS multiple reaction monitoring method. The average intraassay and interassay variation values for PK, MK-4, and MK-7 were <10%. Average spiked recoveries from authentic compounds added to normal human plasma samples for PK, MK-4, and MK-7 were 98-102%. Mean plasma concentrations of PK, MK-4, and MK-7 from healthy subjects (n = 20) were 1.22 +/-0.57, 0.39 +/- 0.46, and 6.37 +/- 7.45 ng/mL, respectively. We conclude that this novel LC-APCI-MS/MS method should be useful for the evaluation of vitamin K status in postmenopausal women and elderly subjects and provides useful information for the treatment and prevention of osteoporosis with vitamin K.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call