Abstract

The noise power spectrum (NPS) in computed tomography (CT) images potentially varies with the X-ray tube angle in a spiral orbit of the helical scan. The purpose of this study was to propose a method for measuring the NPS for each angle of the X-ray tube. Images of the water phantom were acquired using a helical scan. As a conventional method, we measured the two-dimensional (2D) NPS from each image and averaged them; the obtained 2D-NPS was referred to as NPSconventional. In the proposed method, we made the X-ray tube angle θ (0°≤θ<360°) to correspond to the image according to each slice position of the images that located within the travel distance of the CT scan table per 360° rotation of the X-ray tube. We obtained the 2D-NPS from each image and assigned the θ (0°, 30°, 60°, 90°, 120°, 150°, 180°); the obtained 2D-NPS was referred to as NPSsθ. The NPSsθ was compared to the NPSconventional. Also, we investigated the dependency of the NPSsθ on the θ. The NPSconventional was found to be isotropic, and in contrast, the NPSsθ was anisotropic. The NPSsθ showed a continuously rotational change while increasing the θ. There was an excellent correlation (R2>0.999) between the rotation angle of NPSθ and the θ. The proposed method was demonstrated to be effective for evaluating anisotropic noise characteristics depending on the X-ray tube angle.