Abstract
To measure the latency between human motion stimulation and stereo image display response in a visual feedback-based minimally invasive surgical (MIS) robotic system, a method was proposed by comparing the orientations of input and output events through image-processing technology. This method used a black bar to keep pace with the measured joint rotating at a number of speeds. During tests, an external camera was placed in front of the apparatus with a proper visual field, so that it can simultaneously view orientations of both bars fixed on the corresponding joints. After quantitatively analyzing the accuracy of the proposed measurement method, the method was applied to a visual feedback-based master–slave robotic system with two-degrees-of-freedom. Experimental results show that the latency of the overall system was approximately 250 ms, and the opposite clearance of the measured joint was in the range of 1.7°–1.9°.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.