Abstract

In order to study the possibility of improving the timing performance of the time of flight (TOF) systems, which are made of plastic scintillator counters, and read out by photomultiplier tubes (PMT) with mesh dynodes and conventional electronics, we have conducted a study using faster PMTs and ultra fast waveform digitizers to read out the plastic scintillators. Different waveform analysis methods are used to calculate the time resolution of such a system. Results are compared with the conventional discriminating method based on a threshold and pulse height. Our tests and analysis show that significant timing performance improvements can be achieved by using this new system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.