Abstract

Functional magnetic resonance imaging (fMRI) is a quickly developing method for non-invasive dynamic brain studies. It uses swift acquisition sequences like echo planar imaging (EPI) that are very sensitive to susceptibility artifacts. These artifacts relate to magnetic field inhomogeneities and may cause geometric distortions. Many methods for correcting these distortions are currently used. Most common are the field mapping methods that use the map of field strength. To create a field map, different approaches can be used and different data must be acquired for each method. This paper compares a commonly used conventional gradient echo (GE) field mapping method with a 3EPI phase images based method. Although the EPI method is more sophisticated and was expected to perform better, the GE field maps showed better results in distortion correction. The cause of this is not in the method’s principle itself, but in its high requirements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.