Abstract
AbstractThe crystal grain size can be quantitatively calculated by Scherrer equation according to the diffraction peak broadening in the XRD curves. Actually, the results calculated by the Scherrer equation are the thickness that perpendicular to the crystal planes. However, in the actual XRD measurements, the broadening of the diffraction peaks is not only because of the Micro‐level changes of crystal such as grain size and lattice distortion, but also due to the instrumental broadening. Thus, the Scherrer equation is less reliable if the full width at half maximum caused by the physical broadening is smaller than that caused by the instrumental broadening. In this paper, it is concluded that the applicable range of the Scherrer equation will increases with the increasing diffraction angle. As an example of Scherrer equation's application, the calculation result for the maximum applicable scope of Si(100) films is 137 nm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.