Abstract

<p><span>Trans Aortic Valve Replacement (TAVR) has offered the cardiology sector of health a new alternative to open heart surgeries which treat aortic stenosis. The technologies used by TAVR manufacturers are kept private. Our research goal was to develop a process that allows college level laboratories to fabricate their own TAVR stents in order to research new designs and methods of fabrication which may improve current TAVR practices. By creating a solid model of a stent cell design in SolidWorks, we were able to export a cutting pattern we used with a waterjet. The stent frame was then hand polished to prepare for fabric skirting and leaflet attachment. Synthetic ripstop fabric was cut using a commercial fabric cutting machine and attached to the frame using a waterproof glue. Future research entails welding techniques, improved polishing methods, and implantation into a mechanical system. This prototype could be used for TAVR related research and surgical training simulations.</span></p>

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.