A computer-controlled system, included a sequential injection system and an electrochemical detector, was developed for the determination of mercury(II) ion (Hg2+) concentrations. The screen-printed carbon electrode (SPCE), a well-known low-cost electrode, was modified by overlaying in situ with a gold film (Au-SPCE). Square wave anodic stripping voltammetry (SWASV) using the Au-SPCE was performed with a two-step deposition potential of an initial −0.5V vs Ag/AgCl for the preparation of the gold film and then a deposition potential of +0.2V vs Ag/AgCl for the preconcentration of mercury. An on-line medium exchange method was included in the injection step to reduce the effect of chloride ions, whilst a solid phase extraction cartridge was used to remove interference for copper(II) ions. The flow pattern of the sequentially injected solutions with the synchronized two-step deposition potentials did not significantly affect the detection of Hg2+. Square wave parameters of a 4mV step potential, 150Hz frequency and 30mV amplitude gave an optimal Hg2+ detection sensitivity with a limit of detection of 0.22μgL−1 (sample volume of 0.9mL) without any significant interference effect. This method was successfully applied to determine Hg2+ concentrations in real water samples, including in the chloride-rich samples of sea water and table salt, with very good accuracy (recovery in the range of 96.0–101%).