Abstract

Supercritical fluid chromatography (SFC) is drawing considerable interest as separation technique in the pharmaceutical industry. The technique is already well established in chiral separations both analytically and on a preparative scale. The use of SFC as a technique for drug impurity profiling is examined here. To define starting conditions in method development for drug impurity profiling, a set of dissimilar stationary phases is screened in parallel. The possibility to select a set of dissimilar columns using the retention factors (k-values) for a set of 64 drugs measured on 27 columns in SFC was examined. Experiments were carried out at a back-pressure of 150bar and 25°C with a mobile phase consisting of CO2 and methanol with 0.1% isopropylamine (5–40% over 10min) at a flow rate of 3mL/min. These k-values were then used to calculate correlation coefficients on the one hand and to perform a principal component analysis on the other. The Kennard and Stone algorithm, besides dendrograms and correlation-coefficient colour maps were used to select a set of 6 dissimilar stationary phases. The stationary phase characterization results from this study were compared to those from previous studies found in the literature. Retention mechanisms for compounds possessing different properties were also evaluated. The dissimilarity of the selected subset of 6 stationary phases was validated using mixtures of compounds with similar properties and structures, as one can expect in a drug impurity profile.