Abstract

A simple, specific and highly sensitive assay using liquid chromatography-tandem mass spectrometry with electrospray ionization was developed for the determination of olprinone in human plasma. Following the addition of codeine as an internal standard and simple liquid-liquid extraction with ethyl acetate-isopropanol (95:5, v/v), the analytes were separated on a Zorbax SB-C18 column (4.6 × 150 mm i.d., 5 µm) and eluted with a mobile phase consisting of methanol-10 mmol/L ammonium acetate containing 1% formic acid (50:50, v/v) at a flow rate of 1.0 mL/min. Detection was achieved with electrospray positive ionization mass spectrometry using multiple reaction monitoring mode. Linear calibration curves were obtained in the concentration range of 0.1-100 ng/mL (r = 0.9995), with a lower limit of quantification of 0.1 ng/mL. The intra-day and inter-day precision values, as relative deviation, were below 6.06% and the accuracy, as relative error, was below 9.92% at all quality control concentrations. The method was applicable to clinical pharmacokinetic study of an olprinone hydrochloride hydrate injection in healthy Chinese volunteers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.